2-1. Troubleshooting
2-1-1. ICP-6520/6530/3000
The status indication lamp A/B is not lit in green
Flow1
The status indication lamp
A/B is not lit in green.
YES
Is the internal power switch
(A/B) turned on?
YES
Is the power cord connected
correctly?
YES
The possible cause is the
power unit failure.
The unit does not start (POWER switch is not lit)
Flow2
The POWER switch is not lit.
YES
NO
Is the status indication lamp
lit?
YES
The possible cause is the
failure of this unit.
ICP-6520/6530/3000/6511
Service Overview
NO
Turn on the power switch
(A/B).
NO
Connect the power cord
correctly.
Check the item in Flow1
"The status indication lamp
A/B is not lit in green".
Section 2
Switched and indicators, etc. on the panel are not lit
Flow3
Switched and indicators, etc.
on the panel are not lit.
YES
Is the status indication lamp
lit?
YES
Is the POWER switch lit?
YES
Is the MVS LAN cable
connected to the connector
(MVS) of this unit?
YES
The possible cause is the
failure of this unit.
2-1-2. ICP-6511
No image is displayed on the menu screen.
Flow4
No image is displayed on
the menu screen.
YES
Is the AC adapter connected
correctly?
YES
Is the connecting cable
connected correctly?
YES
Is the ICP-6511 switch lit?
YES
The possible cause is the
failure of the LCD panel or
drive circuit of ICP-6511.
Check the item in Flow1
NO
"The status indication lamp
A/B is not lit in green".
Check the item in Flow2
NO
"The unit does not start.
(POWER switch is not lit)".
NO
Connect the LAN cable.
NO
Connect the AC adapter.
Connect the DVI cable and
NO
USB cable with
ICP-6520/6530/3000.
NO
The possible cause is the
failure of ICP-6511.
2-1 (E)