Valor 2000 Series Scales EN-1 1. INTRODUCTION This manual contains installation, operation and maintenance instructions for the Valor 2000 Series Scales. Please read the manual completely before using the scale. 1.1 Safety Precautions Please follow these safety precautions: • Verify that the AC Adapter input voltage matches the local AC power supply.
EN-2 Valor 2000 Series Scales TABLE 1-1. CONTROL FUNCTIONS. Button Primary Function ON/ 0/T UNITS (Short Press) Turns scale ON. Changes the weighing Unit. If scale is On: Sets zero. Performs a tare operation. Secondary Menu Function Turns scale off.
Page 5
Valor 2000 Series Scales EN-3 Level Bubble Adjustable Feet Figure 2-1. Leveling the Scale. IMPORTANT: To relieve any internal pressure and allow proper airflow inside the scale, unscrew the two rear leveling feet slightly, then level the scale accordingly. When cleaning the scale, the rear leveling feet must be screwed in tightly to avoid entry of water inside the scale.
• Install the rubber seal securely to ensure the scale is protected when operating in a wet environment. CAUTION: Battery is to be replaced only by an authorized Ohaus service dealer. Risk of explosion can occur if replaced with the wrong type or connected improperly.
Valor 2000 Series Scales EN-5 2.5 Initial Calibration When the scale is operated for the first time, a Span calibration is recommended to ensure accurate weighing results. Before performing the calibration, be sure to have the appropriate calibration weight. Refer to section 4.2 for Span calibration procedures.
EN-6 Valor 2000 Series Scales 4.2 CAL Menu SPAN Perform Span calibration Yes or No. Perform Linearity calibration Yes or No. NOTE: • Ensure the appropriate calibration masses are available before beginning calibration (see Table 4-1). • Ensure that the scale is level and stable during the entire calibration process.
Valor 2000 Series Scales EN-7 The display flashes the span calibration point. Place the specified weight on the scale and press the Yes button. To choose a different span point, repeatedly press the No button to increment the selections. When the desired value is displayed, place the specified weight on the scale and press the Yes button.
EN-8 Valor 2000 Series Scales The display flashes the full calibration point. Place the specified weight on the scale and press the Yes button. The display shows --C-- while the full calibration point is established. If linearity calibration was successful, dONE is displayed and the scale exits to the active weighing mode and displays the actual weight value.
Valor 2000 Series Scales EN-9 4.5 A OFF Menu AUTO SHUT OFF Set the automatic shut off to 5, 10, 15 minutes (shut off after 5 minutes, etc.) or OFF (scale stays on). 4.6 FILTR Menu FILTERING LEVEL Set the filter level to LO (low), MEd (medium), or HI (high).
EN-10 Valor 2000 Series Scales Lock Switch Figure 5-1. Lock Switch. When accessing the bottom of the scale, avoid placing the scale upside down on the pan or sub-platform. Place the scale on its side. 5.2 Verification Before this product can be used in a trade approved application, it must be inspected in accordance with local weights and measures regulations.
Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panels. 6.2 Troubleshooting The following table lists common problems and possible causes and remedies. If the problem persists, contact Ohaus or your authorized dealer. TABLE 6-1. TROUBLESHOOTING. Symptom Possible Cause(s)
6.3 Service Information If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. For service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8.00 AM and 5:00 PM EST. An Ohaus product service specialist will be available to provide assistance.
Valor 2000 Series Scales EN-13 7.2. Drawings and Dimensions 249 mm (9.8 in.) 229 mm 233 mm (9.0 in.) (9.2 in.) 184 mm (7.2 in.) 142 mm (5.6 in.) Figure 7-1. Valor 2000 Series Overall Dimensions.
Page 16
Mark Standard This product conforms to the EMC directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. The complete declaration of Conformity is available from Ohaus Corporation AS/NZS4251.1 Emission; AS/NZS4252.1 Immunity NSF/ANSI 169 – 2005 NSF/ANSI /3-A 14159-1 - 2002 Important notice for verified weighing instruments...
Page 17
Cet appareil numérique de la classe B est conforme à la norme NMB-003 du Canada. ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements.
During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.